Recall Z-2609-2026— Medline Industries, LP
Class II · Ongoing
Class II FDA medical device recall by Medline Industries, LP, initiated 2026-05-28. FDA's recall record names no specific 510(k) clearance for this event. Product: Medline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG DYNJ905855 KIT UNIVERSITY CABG PACK DYNJ905857 OH MED RN CDS DYNJ905936B. Reason: Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss..
- Recalling firm
- Medline Industries, LP
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-05-28
- Firm location
- Northfield, IL, United States
Product description
Medline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG DYNJ905855 KIT UNIVERSITY CABG PACK DYNJ905857 OH MED RN CDS DYNJ905936B
Reason for recall
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Data sourced from openFDA. This site is unofficial and independent of the FDA.