Recall Z-2609-2016— Baxter Healthcare Corp.

Class II · Terminated

Class II FDA medical device recall by Baxter Healthcare Corp., initiated 2013-10-07, terminated 2016-12-23. It names one FDA 510(k) clearance: K102936. Product: 1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. 2) The HomeChoice PRO automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. The HomeChoice/HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.. Reason: In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop, but cannot bypass, the active Initial Drain. This is to mitigate against Unintended Increased Intraperitoneal Volume (IIVP). This can cause serious problems in patients with unrelated a co-morbid condition of ascites..

Recalling firm
Baxter Healthcare Corp.
Classification
Class II
Status
Terminated
Initiated
2013-10-07
Terminated
2016-12-23
Firm location
Deerfield, IL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. 2) The HomeChoice PRO automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. The HomeChoice/HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Reason for recall

In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop, but cannot bypass, the active Initial Drain. This is to mitigate against Unintended Increased Intraperitoneal Volume (IIVP). This can cause serious problems in patients with unrelated a co-morbid condition of ascites.