Recall Z-2608-2026— Medline Industries, LP

Class II · Ongoing

Class II FDA medical device recall by Medline Industries, LP, initiated 2026-05-28. FDA's recall record names no specific 510(k) clearance for this event. Product: Medline Convenience kits containing recalled Swan-Ganz Catheters CENTRAL AND ARTERIAL LINE TOTE DYNJ8407. Reason: Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss..

Recalling firm
Medline Industries, LP
Classification
Class II
Status
Ongoing
Initiated
2026-05-28
Firm location
Northfield, IL, United States

Product description

Medline Convenience kits containing recalled Swan-Ganz Catheters CENTRAL AND ARTERIAL LINE TOTE DYNJ8407

Reason for recall

Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.

Data sourced from openFDA. This site is unofficial and independent of the FDA.