Recall Z-2608-2020— Skeletal Kinetics, Llc
Class II · Terminated
Class II FDA medical device recall by Skeletal Kinetics, Llc, initiated 2020-04-06, terminated 2021-02-24. It names one FDA 510(k) clearance: K100986. Product: EXACTECH OSSILIX FORM (MX), 10CC. Product Number: 660-01-10, UDI: 813845021037 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.. Reason: Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion..
- Recalling firm
- Skeletal Kinetics, Llc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-04-06
- Terminated
- 2021-02-24
- Firm location
- Addison, TX, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
EXACTECH OSSILIX FORM (MX), 10CC. Product Number: 660-01-10, UDI: 813845021037 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
Reason for recall
Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.