Class I FDA medical device recall by Insulet Corporation, initiated 2026-05-20. It names one FDA 510(k) clearance: K211575. Product: Omnipod (Eros) Pod. Includes the below Model/REF Numbers: 1. POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). 2. ZXR425 (10-Pack), ZXR420 (Single Pod). 3. ZXP425 (10 Pack), ZXP420 (Single Pod).. Reason: External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death..
- Recalling firm
- Insulet Corporation
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2026-05-20
- Firm location
- Acton, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Omnipod (Eros) Pod. Includes the below Model/REF Numbers: 1. POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). 2. ZXR425 (10-Pack), ZXR420 (Single Pod). 3. ZXP425 (10 Pack), ZXP420 (Single Pod).
Reason for recall
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.