Recall Z-2595-2023— MEDLINE INDUSTRIES, LP - Northfield
Class II · Ongoing
Class II FDA medical device recall by MEDLINE INDUSTRIES, LP - Northfield, initiated 2023-08-03. FDA's recall record names no specific 510(k) clearance for this event. Product: MEDLINE STERILE IRIS SCISSORS CVD/STD, REF DYNJ04049. Reason: 4.5" Iris Scissors, Sterile, Curved (DYNJ04049) are being recalled due to the potential of the tip protector to fall off. The sharp tip of the scissors could compromise the sterile barrier system by puncturing large and/or small holes that may not be visible to the user..
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-08-03
- Firm location
- Northfield, IL, United States
Product description
MEDLINE STERILE IRIS SCISSORS CVD/STD, REF DYNJ04049
Reason for recall
4.5" Iris Scissors, Sterile, Curved (DYNJ04049) are being recalled due to the potential of the tip protector to fall off. The sharp tip of the scissors could compromise the sterile barrier system by puncturing large and/or small holes that may not be visible to the user.