Recall Z-2592-2020— Skeletal Kinetics, Llc

Class II · Terminated

Class II FDA medical device recall by Skeletal Kinetics, Llc, initiated 2020-04-06, terminated 2021-02-24. It names one FDA 510(k) clearance: K051123. Product: ACUMED CALLOS INJECT 10CC. Product Number: 65-0010-S, UDI: 813845020245 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.. Reason: Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion..

Recalling firm
Skeletal Kinetics, Llc
Classification
Class II
Status
Terminated
Initiated
2020-04-06
Terminated
2021-02-24
Firm location
Addison, TX, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ACUMED CALLOS INJECT 10CC. Product Number: 65-0010-S, UDI: 813845020245 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.

Reason for recall

Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.