Recall Z-2592-2017— Ion Beam Applications S.A.
Class II · Terminated
Class II FDA medical device recall by Ion Beam Applications S.A., initiated 2017-02-09, terminated 2023-07-14. It names one FDA 510(k) clearance: K152224. Product: Proteus 235. Reason: It has been observed during testing that there could be clinically non-negligible amount of dark current (a well-known physics effect observed in cyclotron) in Pencil Beam Scanning (PBS) treatment mode when system is not irradiating between layers..
- Recalling firm
- Ion Beam Applications S.A.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-02-09
- Terminated
- 2023-07-14
- Firm location
- Louvain La Neuve, Belgium
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Proteus 235
Reason for recall
It has been observed during testing that there could be clinically non-negligible amount of dark current (a well-known physics effect observed in cyclotron) in Pencil Beam Scanning (PBS) treatment mode when system is not irradiating between layers.