Recall Z-2591-2017— Merge Healthcare, Inc.
Class II · Terminated
Class II FDA medical device recall by Merge Healthcare, Inc., initiated 2017-05-08, terminated 2019-08-08. It names one FDA 510(k) clearance: K082144. Product: Merge PACS software. Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication, and storage of studies produced by digital modalities, including digital mammography. Reason: When cut lines are shown on an image, the lines may go from being vertical and then shift to horizontal causing confusion to the physician. This may cause a misdiagnosis..
- Recalling firm
- Merge Healthcare, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-05-08
- Terminated
- 2019-08-08
- Firm location
- Hartland, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Merge PACS software. Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication, and storage of studies produced by digital modalities, including digital mammography
Reason for recall
When cut lines are shown on an image, the lines may go from being vertical and then shift to horizontal causing confusion to the physician. This may cause a misdiagnosis.