Recall Z-2581-2021— Biodex Medical Systems, Inc.

Class II · Terminated

Class II FDA medical device recall by Biodex Medical Systems, Inc., initiated 2021-08-16, terminated 2024-05-30. It names one FDA 510(k) clearance: K090296. Product: AtomLab 500Plus Dose CalibratorSoftware 2.0.00 through Revision 2.0.08, inclusive Model: 086-332. Reason: When deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have incorrect dial settings and may lead to an incorrect dose measurement and/or misadministration.

Recalling firm
Biodex Medical Systems, Inc.
Classification
Class II
Status
Terminated
Initiated
2021-08-16
Terminated
2024-05-30
Firm location
Shirley, NY, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

AtomLab 500Plus Dose CalibratorSoftware 2.0.00 through Revision 2.0.08, inclusive Model: 086-332

Reason for recall

When deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have incorrect dial settings and may lead to an incorrect dose measurement and/or misadministration