Recall Z-2576-2024— Merit Medical Systems, Inc.

Class II · Ongoing

Class II FDA medical device recall by Merit Medical Systems, Inc., initiated 2024-06-11. FDA's recall record names no specific 510(k) clearance for this event. Product: Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Safety Paracentesis Procedure Tray (Paracentesis Tray kit), REF: SPPT-100/D SPPT-5F-10/D SPPT-5F-10L/B SPPT-5F-10S SPPT-5F-12PL SPPT-5F-17PL SPPT-5F-7/D SPPT-5F-7L/B SPPT-5F-7S SPPT-6F-10L SPPT-6F-10S SPPT-6F-10S SPPT-6F-12PL SPPT-6F-17PL SPPT-8F-10L SPPT-8F-12PL TAPS-100 TAPS-5F-10LT TAPS-5F-10T TAPS-5F-12PLT TAPS-5F-17PLT TAPS-5F-7LT TAPS-5F-7T TAPS-6F-10LT The Safety Paracentesis Procedure Tray is intended for drainage of fluid and specimen collection for diagnostic purposes or to remove excess fluid collections.. Reason: Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health..

Recalling firm
Merit Medical Systems, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-06-11
Firm location
South Jordan, UT, United States

Product description

Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Safety Paracentesis Procedure Tray (Paracentesis Tray kit), REF: SPPT-100/D SPPT-5F-10/D SPPT-5F-10L/B SPPT-5F-10S SPPT-5F-12PL SPPT-5F-17PL SPPT-5F-7/D SPPT-5F-7L/B SPPT-5F-7S SPPT-6F-10L SPPT-6F-10S SPPT-6F-10S SPPT-6F-12PL SPPT-6F-17PL SPPT-8F-10L SPPT-8F-12PL TAPS-100 TAPS-5F-10LT TAPS-5F-10T TAPS-5F-12PLT TAPS-5F-17PLT TAPS-5F-7LT TAPS-5F-7T TAPS-6F-10LT The Safety Paracentesis Procedure Tray is intended for drainage of fluid and specimen collection for diagnostic purposes or to remove excess fluid collections.

Reason for recall

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

Data sourced from openFDA. This site is unofficial and independent of the FDA.