Recall Z-2574-2014— Elekta, Inc.

Class II · Terminated

Class II FDA medical device recall by Elekta, Inc., initiated 2014-08-18, terminated 2017-08-08. It names one FDA 510(k) clearance: K013112. Product: FOCAL Sim Workstation, Radiation Treatment Planning System. Used for patient contouring and viewing of treatment plan outputs. Also able to perform CT simulations and forward their information to the RTP System for dose calculation.. Reason: When DICOM is exporting an arc plan (VMAT, Dynamic Conformal of 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not match the approved treatment plan..

Recalling firm
Elekta, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-08-18
Terminated
2017-08-08
Firm location
Atlanta, GA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

FOCAL Sim Workstation, Radiation Treatment Planning System. Used for patient contouring and viewing of treatment plan outputs. Also able to perform CT simulations and forward their information to the RTP System for dose calculation.

Reason for recall

When DICOM is exporting an arc plan (VMAT, Dynamic Conformal of 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not match the approved treatment plan.