Recall Z-2573-2023— Landauer
Class II · Completed
Class II FDA medical device recall by Landauer, initiated 2023-07-12. It names one FDA 510(k) clearance: K192196. Product: NanoDots, Model Numbers: a) BC30023, description: QC Set For Microstar Dots; b) BC30083, description: nanoDot QC Set - 80kVp - Sale; c) BC30084, description: nanoDot QC Set - 80kVp - No Charge; d) BC30088, description: nanoDot Calibration Set - 80kVp - No Charge; e) BC30095, description: Dot, D1DNN, in carrier sealed in bag; f) BC30110, description: Calibration set, nanoDot; g) BC30141, description: nanoDot QC Set - Cs137 - Sale; h) BC30142, description: nanoDot Calibration set - Cs137 - Sale; radiation monitoring dosimeter used with the microSTAR readers. Reason: LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure..
- Recalling firm
- Landauer
- Classification
- Class II
- Status
- Completed
- Initiated
- 2023-07-12
- Firm location
- Glenwood, IL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
NanoDots, Model Numbers: a) BC30023, description: QC Set For Microstar Dots; b) BC30083, description: nanoDot QC Set - 80kVp - Sale; c) BC30084, description: nanoDot QC Set - 80kVp - No Charge; d) BC30088, description: nanoDot Calibration Set - 80kVp - No Charge; e) BC30095, description: Dot, D1DNN, in carrier sealed in bag; f) BC30110, description: Calibration set, nanoDot; g) BC30141, description: nanoDot QC Set - Cs137 - Sale; h) BC30142, description: nanoDot Calibration set - Cs137 - Sale; radiation monitoring dosimeter used with the microSTAR readers
Reason for recall
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.