Recall Z-2560-2018— Galt Medical Corporation

Class II · Terminated

Class II FDA medical device recall by Galt Medical Corporation, initiated 2018-05-02, terminated 2020-07-20. FDA's recall record names no specific 510(k) clearance for this event. Product: 1) Drainer Centesis Catheter, Sterile EO, Rx Only, Catalog #: a) DRC-001-05, b) DRC-001-06, c) DRC-001-15, 2) Centeze Centesis Catheter, Sterile EO, Rx Only, Catalog #: a) DRC-002-02, b) DRC-002-05, c) DRC-002-06, d) DRC-002-08,. Reason: The products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce fever, chills, and hypotension in a patient..

Recalling firm
Galt Medical Corporation
Classification
Class II
Status
Terminated
Initiated
2018-05-02
Terminated
2020-07-20
Firm location
Garland, TX, United States

Product description

1) Drainer Centesis Catheter, Sterile EO, Rx Only, Catalog #: a) DRC-001-05, b) DRC-001-06, c) DRC-001-15, 2) Centeze Centesis Catheter, Sterile EO, Rx Only, Catalog #: a) DRC-002-02, b) DRC-002-05, c) DRC-002-06, d) DRC-002-08,

Reason for recall

The products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce fever, chills, and hypotension in a patient.

Data sourced from openFDA. This site is unofficial and independent of the FDA.