Class II FDA medical device recall by Roche Diagnostics Hematology, initiated 2018-05-07, terminated 2020-05-12. It names one FDA 510(k) clearance: K171655. Product: cobas m 511 integrated hematology analyzer, Model Number 07261691190 Product Usage: The cobas m 511 integrated hematology analyzer is a quantitative, automated analyzer with cell locating capability. It is intended for in vitro diagnostic use by a skilled operator in the clinical laboratory. The system prepares a stained microscope slide from EDTA anticoagulated whole blood. It utilizes computer imaging to count the formed elements of blood and provide an image-based assessment of cell morphology, which may be reviewed by the operator, and also allows for manual classification of unclassified cells. The instrument reports the following parameters: RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, %NRBC, #NRBC, WBC, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EO, #EO, %BASO, #BASO, PLT, MPV, %RET, #RET, HGB-RET.. Reason: Discrepant results have been reported in patients with severe microcytic anemia (e.g.iron deficiency, thalassemia) and the following parameters are affected: RBC (red blood cell count), HGB (hemoglobin concentration), MCH (mean corpuscular hemoglobin), HCT (hematocrit), and MCV (mean corpuscular volume)..
- Recalling firm
- Roche Diagnostics Hematology
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-05-07
- Terminated
- 2020-05-12
- Firm location
- Westborough, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
cobas m 511 integrated hematology analyzer, Model Number 07261691190 Product Usage: The cobas m 511 integrated hematology analyzer is a quantitative, automated analyzer with cell locating capability. It is intended for in vitro diagnostic use by a skilled operator in the clinical laboratory. The system prepares a stained microscope slide from EDTA anticoagulated whole blood. It utilizes computer imaging to count the formed elements of blood and provide an image-based assessment of cell morphology, which may be reviewed by the operator, and also allows for manual classification of unclassified cells. The instrument reports the following parameters: RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, %NRBC, #NRBC, WBC, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EO, #EO, %BASO, #BASO, PLT, MPV, %RET, #RET, HGB-RET.
Reason for recall
Discrepant results have been reported in patients with severe microcytic anemia (e.g.iron deficiency, thalassemia) and the following parameters are affected: RBC (red blood cell count), HGB (hemoglobin concentration), MCH (mean corpuscular hemoglobin), HCT (hematocrit), and MCV (mean corpuscular volume).