Recall Z-2554-2025— SEASPINE ORTHOPEDICS CORPORATION

Class II · Ongoing

Class II FDA medical device recall by SEASPINE ORTHOPEDICS CORPORATION, initiated 2025-08-06. It names one FDA 510(k) clearance: K192132. Product: Reef TA Inserter, Part Number: TA2-001003, part of the Reef Interbody System. Reason: Inserter, used as part of a system with an intervertebral body fusion device, may contribute to difficulty disengaging the implant or improper implant positioning, which may may lead to prolonged surgery, intraoperative revision, and may require the use of alternate instrumentation. Devices may also have incorrect UDI-DI marking..

Recalling firm
SEASPINE ORTHOPEDICS CORPORATION
Classification
Class II
Status
Ongoing
Initiated
2025-08-06
Firm location
Carlsbad, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Reef TA Inserter, Part Number: TA2-001003, part of the Reef Interbody System

Reason for recall

Inserter, used as part of a system with an intervertebral body fusion device, may contribute to difficulty disengaging the implant or improper implant positioning, which may may lead to prolonged surgery, intraoperative revision, and may require the use of alternate instrumentation. Devices may also have incorrect UDI-DI marking.