Recall Z-2549-2023— Landauer

Class II · Completed

Class II FDA medical device recall by Landauer, initiated 2023-07-12. It names one FDA 510(k) clearance: K192196. Product: Calibrate (80 kVp) screened nanoDot D2DNS, Model Number 03064-OTO; radiation monitoring dosimeter used with the microSTAR readers. Reason: LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure..

Recalling firm
Landauer
Classification
Class II
Status
Completed
Initiated
2023-07-12
Firm location
Glenwood, IL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Calibrate (80 kVp) screened nanoDot D2DNS, Model Number 03064-OTO; radiation monitoring dosimeter used with the microSTAR readers

Reason for recall

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

Data sourced from openFDA. This site is unofficial and independent of the FDA.