Recall Z-2548-2026— Baxter Healthcare Corporation
Class I · Ongoing
Class I FDA medical device recall by Baxter Healthcare Corporation, initiated 2026-05-21. It names 2 FDA 510(k) clearances: K200988, K192143. Product: Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device. Reason: There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup..
- Recalling firm
- Baxter Healthcare Corporation
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2026-05-21
- Firm location
- Deerfield, IL, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device
Reason for recall
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Data sourced from openFDA. This site is unofficial and independent of the FDA.