Recall Z-2548-2020— DiaSorin Molecular LLC

Class II · Terminated

Class II FDA medical device recall by DiaSorin Molecular LLC, initiated 2020-05-12, terminated 2024-04-19. It names one FDA 510(k) clearance: K190219. Product: MOL3650 Simplexa VZV Direct - Product Usage: is intended for use on the LIAISON MDX instrument for the qualitative detection of varicella-zoster virus (VZV) DNA in cerebrospinal fluid (CSF) from patients with signs and symptoms of VZV infections of the central nervous system (CNS).. Reason: Due to a software malfunction in the Multi Assay Suite, there is the probability of increased occurrence of false negative results..

Recalling firm
DiaSorin Molecular LLC
Classification
Class II
Status
Terminated
Initiated
2020-05-12
Terminated
2024-04-19
Firm location
Cypress, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

MOL3650 Simplexa VZV Direct - Product Usage: is intended for use on the LIAISON MDX instrument for the qualitative detection of varicella-zoster virus (VZV) DNA in cerebrospinal fluid (CSF) from patients with signs and symptoms of VZV infections of the central nervous system (CNS).

Reason for recall

Due to a software malfunction in the Multi Assay Suite, there is the probability of increased occurrence of false negative results.