Recall Z-2546-2014— Zimmer, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer, Inc., initiated 2014-07-22, terminated 2015-09-15. FDA's recall record names no specific 510(k) clearance for this event. Product: Knee Instrument Fixation Screws: 48MM HEADED SCREW 48MM HEADLESS SCREW MIS HEADED SCREW, 48MM MIS HEADLESS SCREW, 48MM. Reason: Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost..

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-07-22
Terminated
2015-09-15
Firm location
Warsaw, IN, United States

Product description

Knee Instrument Fixation Screws: 48MM HEADED SCREW 48MM HEADLESS SCREW MIS HEADED SCREW, 48MM MIS HEADLESS SCREW, 48MM

Reason for recall

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.