Recall Z-2545-2020— Centurion Medical Products Corporation
Class II · Terminated
Class II FDA medical device recall by Centurion Medical Products Corporation, initiated 2019-06-01, terminated 2021-06-21. FDA's recall record names no specific 510(k) clearance for this event. Product: Centurion DIFFICULT IV KIT containing BD's Vacutainer Plastic Lithium Heparin Tubes (BD Vacutainer-BD Codes 367884 & 367960; Centurion code 45177 (VACUTAINER GREEN TOP) and 84815 (VACUTAINER 3ML LT. GREEN). Centurion Kit Code: IV8860. Reason: False elevation of carboxyhemoglobin (COHb) results in blood samples collected using BD Vacutainer¿ Plastic Lithium Heparin tubes with the IL GEM 4000 instrument..
- Recalling firm
- Centurion Medical Products Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-06-01
- Terminated
- 2021-06-21
- Firm location
- Williamston, MI, United States
Product description
Centurion DIFFICULT IV KIT containing BD's Vacutainer Plastic Lithium Heparin Tubes (BD Vacutainer-BD Codes 367884 & 367960; Centurion code 45177 (VACUTAINER GREEN TOP) and 84815 (VACUTAINER 3ML LT. GREEN). Centurion Kit Code: IV8860
Reason for recall
False elevation of carboxyhemoglobin (COHb) results in blood samples collected using BD Vacutainer¿ Plastic Lithium Heparin tubes with the IL GEM 4000 instrument.