Recall Z-2544-2023— Landauer
Class II · Completed
Class II FDA medical device recall by Landauer, initiated 2023-07-12. It names one FDA 510(k) clearance: K192196. Product: nanoDot D2DNS, Model Numbers: a) 03057-1MO; b) 03057-3MO; c) 03057-OTO; d) 03057-SMO; radiation monitoring dosimeter used with the microSTAR readers. Reason: LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure..
- Recalling firm
- Landauer
- Classification
- Class II
- Status
- Completed
- Initiated
- 2023-07-12
- Firm location
- Glenwood, IL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
nanoDot D2DNS, Model Numbers: a) 03057-1MO; b) 03057-3MO; c) 03057-OTO; d) 03057-SMO; radiation monitoring dosimeter used with the microSTAR readers
Reason for recall
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
Data sourced from openFDA. This site is unofficial and independent of the FDA.