Recall Z-2542-2015— Wright Medical Technology, Inc.
Class II · Terminated
Class II FDA medical device recall by Wright Medical Technology, Inc., initiated 2015-07-21, terminated 2018-08-01. It names one FDA 510(k) clearance: K132594. Product: CrossCHECK Non-Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-N3008; 3.0 x 10 mm, REF CCP-N3010; 3.0 x 12 mm, REF CCP-N3012; 3.0 x 14 mm, REF CCP-N3014l; 3.0 x 16 mm, REF CCP-N3016; 3.0 x 18 mm, REF CCP-N3018; 3.0 x 20 mm, REF CCP-N3020; 3.0 x 22 mm, REF CCP-N3022; 3.0 x 24 mm, REF CCP-N3024; 3.0 x 26 mm, REF CCP-N3026; 3.0 x 28 mm, REF CCP-N3028; 3.0 x 30 mm, REF CCP-N3030; 3.0 x 32 mm, REF CCP-N3032; 3.0 x 34 mm, REF CCP-N3034; 3.0 x 36 mm, REF CCP-N3036; 3.0 x 38 mm, REF CCP-N3038; 3.0 x 40 mm, REF CCP-N3040; 3.5 x 8 mm, REF CCP-N3508; 3.5 x 10 mm, REF CCP-N3510; 3.5 x 12 mm, REF CCP-N3512; 3.5 14 mm, REF CCP-N3514; 3.5 x 16 mm, REF CCP-N3516; 3.5 x 18 mm, REF CCP-N3518; 3.5 x 20 mm, REF CCP-N3520; 3.5 x 22 mm, REF CCP-N3522; 3.5 x 24 mm, REF CCP-N3524; 3.5 x 26 mm, REF CCP-N3526; 3.5 x 28 mm, REF CCP-N3528; 3.5 x 30 mm, REF CCP-N3530; 3.5 x 32 mm, REF CCP-N3532; 3.5 x 34 mm, REF CCP-N3534; 3.5 x 36 mm, REF CCP-N3536; 3.5 x 38 mm, REF CCP-N3538; 3.5 x 40 mm, REF CCP-N3540. The CROSSCHECK Plating System is indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones in the hand, feet wrist and ankles, fingers and toes. The system may be used in both pediatric and adult patients. The device is intended for single use.. Reason: These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction..
- Recalling firm
- Wright Medical Technology, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-07-21
- Terminated
- 2018-08-01
- Firm location
- Memphis, TN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
CrossCHECK Non-Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-N3008; 3.0 x 10 mm, REF CCP-N3010; 3.0 x 12 mm, REF CCP-N3012; 3.0 x 14 mm, REF CCP-N3014l; 3.0 x 16 mm, REF CCP-N3016; 3.0 x 18 mm, REF CCP-N3018; 3.0 x 20 mm, REF CCP-N3020; 3.0 x 22 mm, REF CCP-N3022; 3.0 x 24 mm, REF CCP-N3024; 3.0 x 26 mm, REF CCP-N3026; 3.0 x 28 mm, REF CCP-N3028; 3.0 x 30 mm, REF CCP-N3030; 3.0 x 32 mm, REF CCP-N3032; 3.0 x 34 mm, REF CCP-N3034; 3.0 x 36 mm, REF CCP-N3036; 3.0 x 38 mm, REF CCP-N3038; 3.0 x 40 mm, REF CCP-N3040; 3.5 x 8 mm, REF CCP-N3508; 3.5 x 10 mm, REF CCP-N3510; 3.5 x 12 mm, REF CCP-N3512; 3.5 14 mm, REF CCP-N3514; 3.5 x 16 mm, REF CCP-N3516; 3.5 x 18 mm, REF CCP-N3518; 3.5 x 20 mm, REF CCP-N3520; 3.5 x 22 mm, REF CCP-N3522; 3.5 x 24 mm, REF CCP-N3524; 3.5 x 26 mm, REF CCP-N3526; 3.5 x 28 mm, REF CCP-N3528; 3.5 x 30 mm, REF CCP-N3530; 3.5 x 32 mm, REF CCP-N3532; 3.5 x 34 mm, REF CCP-N3534; 3.5 x 36 mm, REF CCP-N3536; 3.5 x 38 mm, REF CCP-N3538; 3.5 x 40 mm, REF CCP-N3540. The CROSSCHECK Plating System is indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones in the hand, feet wrist and ankles, fingers and toes. The system may be used in both pediatric and adult patients. The device is intended for single use.
Reason for recall
These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction.