Recall Z-2532-2020— Bound Tree Medical

Class II · Terminated

Class II FDA medical device recall by Bound Tree Medical, initiated 2020-05-14, terminated 2023-12-07. FDA's recall record names no specific 510(k) clearance for this event. Product: Curaplex Ambu King LTS-D Supraglottic Airway Kit with ET Tube Part Number: KLTSD404K Curaplex Ambu¿ King LTS-D" Supraglottic Airway Kit with ET Tube (hereafter the Kit which contains all of the components necessary to successfully obtain an open patient airway with an Ambu¿ King LTS-D" Supraglottic Airway.. Reason: Kit contained an incorrect catheter-tip syringe not compatible with the luer lock connector of the included airway. As a result, may be unable to inflate the balloons of the supraglottic airway to create a secure seal in the patient s airway as intended by the kit.

Recalling firm
Bound Tree Medical
Classification
Class II
Status
Terminated
Initiated
2020-05-14
Terminated
2023-12-07
Firm location
Dublin, OH, United States

Product description

Curaplex Ambu King LTS-D Supraglottic Airway Kit with ET Tube Part Number: KLTSD404K Curaplex Ambu¿ King LTS-D" Supraglottic Airway Kit with ET Tube (hereafter the Kit which contains all of the components necessary to successfully obtain an open patient airway with an Ambu¿ King LTS-D" Supraglottic Airway.

Reason for recall

Kit contained an incorrect catheter-tip syringe not compatible with the luer lock connector of the included airway. As a result, may be unable to inflate the balloons of the supraglottic airway to create a secure seal in the patient s airway as intended by the kit