Recall Z-2531-2020— Philips North America, LLC
Class II · Terminated
Class II FDA medical device recall by Philips North America, LLC, initiated 2020-06-08, terminated 2023-01-09. It names one FDA 510(k) clearance: K993383. Product: Philips M3015A Microstream CO2 Extension Model No. 862393 -used with an lntelliVue X3 Patient Monitor (Model No. 867030) or a lntelliVue MMX Multi-Measurement Module (Model No. 867036). Reason: Affected M3015A extensions are incompatible with the lntelliVue X3 Patient Monitor (Model No. 867030) and lntelliVue MMX multi-measurement server (Model No. 867036).If they are connected to a host monitor, communication with the X3 or MMX could be disrupted, resulting in a loss of monitoring on the host monitor..
- Recalling firm
- Philips North America, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-06-08
- Terminated
- 2023-01-09
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Philips M3015A Microstream CO2 Extension Model No. 862393 -used with an lntelliVue X3 Patient Monitor (Model No. 867030) or a lntelliVue MMX Multi-Measurement Module (Model No. 867036)
Reason for recall
Affected M3015A extensions are incompatible with the lntelliVue X3 Patient Monitor (Model No. 867030) and lntelliVue MMX multi-measurement server (Model No. 867036).If they are connected to a host monitor, communication with the X3 or MMX could be disrupted, resulting in a loss of monitoring on the host monitor.