Recall Z-2530-2023— Hamilton Medical, Inc.

Class I · Ongoing

Class I FDA medical device recall by Hamilton Medical, Inc., initiated 2023-07-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator. Reason: Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation..

Recalling firm
Hamilton Medical, Inc.
Classification
Class I
Status
Ongoing
Initiated
2023-07-26
Firm location
Reno, NV, United States

Product description

Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator

Reason for recall

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Data sourced from openFDA. This site is unofficial and independent of the FDA.