Recall Z-2528-2023— Hamilton Medical, Inc.
Class I · Ongoing
Class I FDA medical device recall by Hamilton Medical, Inc., initiated 2023-07-26. It names one FDA 510(k) clearance: K153046. Product: Hamilton-MR1, REF: 161010, Intensive Care Ventilator. Reason: Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation..
- Recalling firm
- Hamilton Medical, Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2023-07-26
- Firm location
- Reno, NV, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Hamilton-MR1, REF: 161010, Intensive Care Ventilator
Reason for recall
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.