Recall Z-2527-2023— Hamilton Medical, Inc.
Class I · Ongoing
Class I FDA medical device recall by Hamilton Medical, Inc., initiated 2023-07-26. It names one FDA 510(k) clearance: K181216. Product: Hamilton-T1, REF: 161009, Intensive Care and Transport Ventilator. Reason: Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation..
- Recalling firm
- Hamilton Medical, Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2023-07-26
- Firm location
- Reno, NV, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Hamilton-T1, REF: 161009, Intensive Care and Transport Ventilator
Reason for recall
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Data sourced from openFDA. This site is unofficial and independent of the FDA.