Recall Z-2525-2016— Siemens Healthcare Diagnostics Inc
Class II · Terminated
Class II FDA medical device recall by Siemens Healthcare Diagnostics Inc, initiated 2016-06-22, terminated 2018-11-29. It names one FDA 510(k) clearance: K984067. Product: Stratus¿ CS STAT Fluorometric Analyzer-microprocessor-controlled instrument that measures certain analytes in body fluids for in vitro diagnostic use. Assays include ~hCG, CRP, CKMB, cTNI, D-Dimer, Myoglobin, and pBNP. SMN: 10444834, 10453531. Reason: Software defect, where either an Above Assay Range or an inaccurate value could potentially be reported, in the Stratus CS Acute Care Diagnostics System regarding TestPak calibration and quality control (QC) status when a TestPak is due to expire within 48 hours..
- Recalling firm
- Siemens Healthcare Diagnostics Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-06-22
- Terminated
- 2018-11-29
- Firm location
- Norwood, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Stratus¿ CS STAT Fluorometric Analyzer-microprocessor-controlled instrument that measures certain analytes in body fluids for in vitro diagnostic use. Assays include ~hCG, CRP, CKMB, cTNI, D-Dimer, Myoglobin, and pBNP. SMN: 10444834, 10453531
Reason for recall
Software defect, where either an Above Assay Range or an inaccurate value could potentially be reported, in the Stratus CS Acute Care Diagnostics System regarding TestPak calibration and quality control (QC) status when a TestPak is due to expire within 48 hours.