Recall Z-2523-2023— Aomori Olympus Co., Ltd.
Class II · Ongoing
Class II FDA medical device recall by Aomori Olympus Co., Ltd., initiated 2023-07-27. It names one FDA 510(k) clearance: K984358. Product: Reusable distal attachments Models: MH-462, MH-463, MH-464, MH-465, MH-466, MH-483, MH-587, MH-588(US), MH-589, MH-590, MH-591, MH-592, MH-593, MH-594, MH-595, MH-596, MH-597, MH-598, MAJ-289, MAJ-290, MAJ-291, MAJ-292, MAJ-293, MAJ-294, MAJ-295, MAJ-296, MAJ-297, MAJ-663, MAJ-664, MAJ-665, MAJ-666. Reason: The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002.
- Recalling firm
- Aomori Olympus Co., Ltd.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-07-27
- Firm location
- Kuroishi, N/A, Japan
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Reusable distal attachments Models: MH-462, MH-463, MH-464, MH-465, MH-466, MH-483, MH-587, MH-588(US), MH-589, MH-590, MH-591, MH-592, MH-593, MH-594, MH-595, MH-596, MH-597, MH-598, MAJ-289, MAJ-290, MAJ-291, MAJ-292, MAJ-293, MAJ-294, MAJ-295, MAJ-296, MAJ-297, MAJ-663, MAJ-664, MAJ-665, MAJ-666
Reason for recall
The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002
Data sourced from openFDA. This site is unofficial and independent of the FDA.