Recall Z-2510-2023— Abbott Medical

Class I · Ongoing

Class I FDA medical device recall by Abbott Medical, initiated 2023-07-18. FDA's recall record names no specific 510(k) clearance for this event. Product: Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensions or leads with up to 16 electrodes total. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG can deliver stimulation with a single program or with multiple programs. Each program can provide stimulation to a single anatomical area or to multiple areas. The IPG communicates wirelessly with system programmers and controllers, and the IPGs are available in small and large sizes to accommodate different power needs. The Proclaim" XR and Proclaim" Plus IPGs are used for Spinal Cord Stimulation (SCS). The Proclaim" DRG IPG is used for Dorsal Root Ganglion (DRG) stimulation. The Infinity" IPG is used for Deep Brain Stimulation (DBS).. Reason: Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators)..

Recalling firm
Abbott Medical
Classification
Class I
Status
Ongoing
Initiated
2023-07-18
Firm location
Plano, TX, United States

Product description

Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensions or leads with up to 16 electrodes total. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG can deliver stimulation with a single program or with multiple programs. Each program can provide stimulation to a single anatomical area or to multiple areas. The IPG communicates wirelessly with system programmers and controllers, and the IPGs are available in small and large sizes to accommodate different power needs. The Proclaim" XR and Proclaim" Plus IPGs are used for Spinal Cord Stimulation (SCS). The Proclaim" DRG IPG is used for Dorsal Root Ganglion (DRG) stimulation. The Infinity" IPG is used for Deep Brain Stimulation (DBS).

Reason for recall

Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).

Data sourced from openFDA. This site is unofficial and independent of the FDA.