Recall Z-2507-2020— Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Class II · Ongoing
Class II FDA medical device recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), initiated 2020-06-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Azure XT DR MRI, Model Numbers: a) W1DR01 and b) W2DR01. Reason: There is a rare communication sequence during the first device interrogation with a SmartSync Device Manager that may result in the temporary suspension of some device features (i.e., battery measurements, Capture Management", Atrial Lead Position Check", EffectivCRT" algorithms, and AdaptivCRT"). This rare interaction results in temporary suspension of automatic threshold testing and output adjustments, and suspension of auto-optimization of CRT therapy. The issue is unlikely to result in clinical impact to the patient, and features are restored upon next programmer device interrogation or presence of a magnet..
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2020-06-04
- Firm location
- Mounds View, MN, United States
Product description
Azure XT DR MRI, Model Numbers: a) W1DR01 and b) W2DR01
Reason for recall
There is a rare communication sequence during the first device interrogation with a SmartSync Device Manager that may result in the temporary suspension of some device features (i.e., battery measurements, Capture Management", Atrial Lead Position Check", EffectivCRT" algorithms, and AdaptivCRT"). This rare interaction results in temporary suspension of automatic threshold testing and output adjustments, and suspension of auto-optimization of CRT therapy. The issue is unlikely to result in clinical impact to the patient, and features are restored upon next programmer device interrogation or presence of a magnet.
Data sourced from openFDA. This site is unofficial and independent of the FDA.