Recall Z-2503-2025— PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Class II · Ongoing
Class II FDA medical device recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., initiated 2025-08-01. It names 5 FDA 510(k) clearances: K163715, K162859, K141979 and 2 more. Product: Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allura Xper FD10/10, Model Number 722005 3. Allura Xper FD20, Model Number 722006 4. Allura Xper FD20 Biplane, Model Number 722008 5. Allura Xper FD10, Model Number 722010 6. Allura Xper FD10/10, Model Number 722011 7. Allura Xper FD20, Model Number 722012 8. Allura Xper FD20 Biplane, Model Number 722013 9. Allura Xper FD20 OR Table, Model Number 722023 10. Allura Xper FD10, Model Number 722026 11. Allura Xper FD10/10, Model Number 722027 12. Allura Xper FD20, Model Number 722028 13. Allura Xper FD20/10, Model Number 722029 14. Allura Xper FD20 OR Table, Model Number 722035 15. Allura Xper FD20/20, Model Number 722038 16. Allura Xper FD20/15, Model Number 722058 17. Allura Xper FD20/15 OR Table, Model Number 722059.. Reason: Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost..
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-08-01
- Firm location
- Best, N/A, Netherlands
510(k) clearances named in this recall
5 on FDA's record- K163715 — Allura Xper R9
- K162859 — Allura Xper FD series and Allura Xper OR Table series
- K141979 — ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES
- K130842 — ALLURA XPER FD OR TABLE SERIES; ALLURA XPER FD10 SERIES; ALLURA XPER FD20 SERIES; ALLURA XPER FD BIPLANE SERIES
- K102005 — ALLURA XPER OR TABLE SERIES
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allura Xper FD10/10, Model Number 722005 3. Allura Xper FD20, Model Number 722006 4. Allura Xper FD20 Biplane, Model Number 722008 5. Allura Xper FD10, Model Number 722010 6. Allura Xper FD10/10, Model Number 722011 7. Allura Xper FD20, Model Number 722012 8. Allura Xper FD20 Biplane, Model Number 722013 9. Allura Xper FD20 OR Table, Model Number 722023 10. Allura Xper FD10, Model Number 722026 11. Allura Xper FD10/10, Model Number 722027 12. Allura Xper FD20, Model Number 722028 13. Allura Xper FD20/10, Model Number 722029 14. Allura Xper FD20 OR Table, Model Number 722035 15. Allura Xper FD20/20, Model Number 722038 16. Allura Xper FD20/15, Model Number 722058 17. Allura Xper FD20/15 OR Table, Model Number 722059.
Reason for recall
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
Data sourced from openFDA. This site is unofficial and independent of the FDA.