Recall Z-2500-2020— XTANT Medical

Class II · Terminated

Class II FDA medical device recall by XTANT Medical, initiated 2020-05-12, terminated 2021-03-19. FDA's recall record names no specific 510(k) clearance for this event. Product: Irix-C Forceps Inserter; Product Code/UDI: T066-0059/ M697T06600591. Reason: The strike plate fractured off from the forceps inserter devices with minimal impaction or force due to unsatisfactory weld..

Recalling firm
XTANT Medical
Classification
Class II
Status
Terminated
Initiated
2020-05-12
Terminated
2021-03-19
Firm location
Belgrade, MT, United States

Product description

Irix-C Forceps Inserter; Product Code/UDI: T066-0059/ M697T06600591

Reason for recall

The strike plate fractured off from the forceps inserter devices with minimal impaction or force due to unsatisfactory weld.

Data sourced from openFDA. This site is unofficial and independent of the FDA.