Recall Z-2500-2018— Philips Healthcare

Class II · Terminated

Class II FDA medical device recall by Philips Healthcare, initiated 2018-06-11, terminated 2024-07-03. FDA's recall record names no specific 510(k) clearance for this event. Product: Centron Product Usage: Vascular,cardiovascular and neurovascular imaging applications,including diagnostic, interventional and minimally invasive procedures. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA,stent placing,atherectomies), pacemaker implantat ions, and electrophysiology (EP). Non-vascular interventions such as drainages,biopsies and vertebroplasties procedures.. Reason: The first time an operator selects a new procedure type during a single examination,the shutter position resets to the open position for the new procedure type. If the shutters had previously been changed during the examination, that setting is not retained after the first time the procedure type is changed during a single examination..

Recalling firm
Philips Healthcare
Classification
Class II
Status
Terminated
Initiated
2018-06-11
Terminated
2024-07-03
Firm location
Andover, MA, United States

Product description

Centron Product Usage: Vascular,cardiovascular and neurovascular imaging applications,including diagnostic, interventional and minimally invasive procedures. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA,stent placing,atherectomies), pacemaker implantat ions, and electrophysiology (EP). Non-vascular interventions such as drainages,biopsies and vertebroplasties procedures.

Reason for recall

The first time an operator selects a new procedure type during a single examination,the shutter position resets to the open position for the new procedure type. If the shutters had previously been changed during the examination, that setting is not retained after the first time the procedure type is changed during a single examination.

Data sourced from openFDA. This site is unofficial and independent of the FDA.