Recall Z-2499-2021— Deroyal Industries, Inc. Lafollette
Class I · Terminated
Class I FDA medical device recall by Deroyal Industries, Inc. Lafollette, initiated 2021-09-02, terminated 2023-12-18. It names one FDA 510(k) clearance: K842648. Product: DeRoyal Procedure Packs containing Smiths Medical NORMOFLO Irrigation Warming Set. Reason: DeRoyal manufactured procedure packs containing a Smiths Medical NORMOFLO Irrigation Warming System which is recalled due to the potential for aluminum ions leaching into warming fluids..
- Recalling firm
- Deroyal Industries, Inc. Lafollette
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2021-09-02
- Terminated
- 2023-12-18
- Firm location
- La Follette, TN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
DeRoyal Procedure Packs containing Smiths Medical NORMOFLO Irrigation Warming Set
Reason for recall
DeRoyal manufactured procedure packs containing a Smiths Medical NORMOFLO Irrigation Warming System which is recalled due to the potential for aluminum ions leaching into warming fluids.