Recall Z-2496-2018— Philips Medical Systems (Cleveland) Inc

Class II · Terminated

Class II FDA medical device recall by Philips Medical Systems (Cleveland) Inc, initiated 2018-05-21, terminated 2020-04-10. It names one FDA 510(k) clearance: K123599. Product: Ingenuity Digital PET/CT System, Model 882446 with the following UDI's: (01)00884838059559(21)900004 (01)00884838059559(21)900005 (01)00884838059559(21)900006 (01)00884838059559(21)900007 (01)00884838059559(21)900008 (01)00884838059559(21)900009 (01)00884838059559(21)900010 (01)00884838059559(21)900011 (01)00884838059559(21)900012 (01)00884838059559(21)900013 (01)00884838059559(21)900014 (01)00884838059559(21)900015 (01)00884838059559(21)900016 (01)00884838059559(21)900017 (01)00884838059559(21)900018 (01)00884838059559(21)900019 (01)00884838059559(21)900020 (01)00884838059559(21)900021 (01)00884838059559(21)900022 (01)00884838059559(21)900023 (01)00884838059559(21)900024 (01)00884838059559(21)900025 (01)00884838059559(21)900026 (01)00884838059559(21)900027 (01)00884838059559(21)900028 (01)00884838059559(21)900029 (01)00884838059559(21)900030 (01)00884838059559(21)900031 (01)00884838059559(21)900032 (01)00884838059559(21)900033 (01)00884838059559(21)900034 (01)00884838059559(21)900035 Product Usage: The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools, including application of the CT scanner for diagnosis and for use in radiation therapy planning.. Reason: The block used to connect the patient pallet to the couch may be assembled incorrectly. A screw may be misaligned in the patient pallet..

Recalling firm
Philips Medical Systems (Cleveland) Inc
Classification
Class II
Status
Terminated
Initiated
2018-05-21
Terminated
2020-04-10
Firm location
Cleveland, OH, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Ingenuity Digital PET/CT System, Model 882446 with the following UDI's: (01)00884838059559(21)900004 (01)00884838059559(21)900005 (01)00884838059559(21)900006 (01)00884838059559(21)900007 (01)00884838059559(21)900008 (01)00884838059559(21)900009 (01)00884838059559(21)900010 (01)00884838059559(21)900011 (01)00884838059559(21)900012 (01)00884838059559(21)900013 (01)00884838059559(21)900014 (01)00884838059559(21)900015 (01)00884838059559(21)900016 (01)00884838059559(21)900017 (01)00884838059559(21)900018 (01)00884838059559(21)900019 (01)00884838059559(21)900020 (01)00884838059559(21)900021 (01)00884838059559(21)900022 (01)00884838059559(21)900023 (01)00884838059559(21)900024 (01)00884838059559(21)900025 (01)00884838059559(21)900026 (01)00884838059559(21)900027 (01)00884838059559(21)900028 (01)00884838059559(21)900029 (01)00884838059559(21)900030 (01)00884838059559(21)900031 (01)00884838059559(21)900032 (01)00884838059559(21)900033 (01)00884838059559(21)900034 (01)00884838059559(21)900035 Product Usage: The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools, including application of the CT scanner for diagnosis and for use in radiation therapy planning.

Reason for recall

The block used to connect the patient pallet to the couch may be assembled incorrectly. A screw may be misaligned in the patient pallet.

Data sourced from openFDA. This site is unofficial and independent of the FDA.