Recall Z-2488-2026— Medline Industries, LP

Class II · Ongoing

Class II FDA medical device recall by Medline Industries, LP, initiated 2026-05-04. FDA's recall record names no specific 510(k) clearance for this event. Product: MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR MMC, Medline Kit Number/SKU DYNJ905368D; 4) KIT PERIVASCULAR MMC, Medline Kit Number/SKU DYNJ905368B; 5) VP SHUNT PACK, Medline Kit Number/SKU DYNJ46530I.. Reason: During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification..

Recalling firm
Medline Industries, LP
Classification
Class II
Status
Ongoing
Initiated
2026-05-04
Firm location
Northfield, IL, United States

Product description

MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR MMC, Medline Kit Number/SKU DYNJ905368D; 4) KIT PERIVASCULAR MMC, Medline Kit Number/SKU DYNJ905368B; 5) VP SHUNT PACK, Medline Kit Number/SKU DYNJ46530I.

Reason for recall

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Data sourced from openFDA. This site is unofficial and independent of the FDA.