Recall Z-2488-2015— Synthes (USA) Products LLC

Class II · Terminated

Class II FDA medical device recall by Synthes (USA) Products LLC, initiated 2015-07-29, terminated 2016-08-18. It names one FDA 510(k) clearance: K142587. Product: VEPTR Vertical Expandable Prosthetic Titanium Rib, inferior cradles also known as caudal rib supports. The VEPTR/VEPTR II devices mechanically stabilize and/or correct thoracic deformities to allow the chest and lungs to grow.. Reason: The last hole on certain VEPTR Inferior Cradles may be partially formed. Fully lengthening the construct to the partially formed last hole may result in mechanical failure of the construct due to rib sleeve breakage, and may require surgury to exchange components. Effects may also include pain, loss of deformity correction, and/or soft tissue/visceral injury..

Recalling firm
Synthes (USA) Products LLC
Classification
Class II
Status
Terminated
Initiated
2015-07-29
Terminated
2016-08-18
Firm location
West Chester, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

VEPTR Vertical Expandable Prosthetic Titanium Rib, inferior cradles also known as caudal rib supports. The VEPTR/VEPTR II devices mechanically stabilize and/or correct thoracic deformities to allow the chest and lungs to grow.

Reason for recall

The last hole on certain VEPTR Inferior Cradles may be partially formed. Fully lengthening the construct to the partially formed last hole may result in mechanical failure of the construct due to rib sleeve breakage, and may require surgury to exchange components. Effects may also include pain, loss of deformity correction, and/or soft tissue/visceral injury.

Data sourced from openFDA. This site is unofficial and independent of the FDA.