Recall Z-2487-2015— Pointe Scientific, Inc.

Class II · Terminated

Class II FDA medical device recall by Pointe Scientific, Inc., initiated 2015-05-19, terminated 2016-05-23. It names one FDA 510(k) clearance: K024006. Product: Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5 ml (2 vials each of Level 1, Level 2 and Level 3). Also sold as individual vials Bottle: Clear glass vials Cap: Rubber seal with colored aluminum seals Glucose-6-phosphate dehydrogenase (G6PD) controls can be used to test for the quantitative and qualitative determination of G6PD in blood.. Reason: A reduction in the reconstituted stability has been identified. Clinicians may notice a drop in recovery of the control after reconstitution, which could result in an out of specification result prior to the current reconstitution claim of 7 days. For this reason we suggest discontinuing use of the control set..

Recalling firm
Pointe Scientific, Inc.
Classification
Class II
Status
Terminated
Initiated
2015-05-19
Terminated
2016-05-23
Firm location
Canton, MI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5 ml (2 vials each of Level 1, Level 2 and Level 3). Also sold as individual vials Bottle: Clear glass vials Cap: Rubber seal with colored aluminum seals Glucose-6-phosphate dehydrogenase (G6PD) controls can be used to test for the quantitative and qualitative determination of G6PD in blood.

Reason for recall

A reduction in the reconstituted stability has been identified. Clinicians may notice a drop in recovery of the control after reconstitution, which could result in an out of specification result prior to the current reconstitution claim of 7 days. For this reason we suggest discontinuing use of the control set.