Recall Z-2486-2025— Reflexion Medical, Inc.
Class II · Ongoing
Class II FDA medical device recall by Reflexion Medical, Inc., initiated 2025-07-22. FDA's recall record names no specific 510(k) clearance for this event. Product: RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008 (01)00860003983812(11)240528(21)X11009 (01)00860003983812(11)240820(21)X11010 (01)00860003983812(11)241107(21)X11011 (01)00860003983812(11)241206(21)X11012 The RefleXion X1 is a radiation therapy system that combines the functionality of an emission computed tomography detection system and a linear accelerator. The device is intended for use with approved fludeoxyglucose F18. The emission computed tomography detection system acquires images of positron-emitting fludeoxyglucose F18 for the purpose of guiding the delivery of megavoltage X-rays for oncologic treatment with radiation therapy using an FDA cleared, authorized, or approved linear accelerator.. Reason: Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potentially lead to adverse events (toxicity)..
- Recalling firm
- Reflexion Medical, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-07-22
- Firm location
- Hayward, CA, United States
Product description
RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008 (01)00860003983812(11)240528(21)X11009 (01)00860003983812(11)240820(21)X11010 (01)00860003983812(11)241107(21)X11011 (01)00860003983812(11)241206(21)X11012 The RefleXion X1 is a radiation therapy system that combines the functionality of an emission computed tomography detection system and a linear accelerator. The device is intended for use with approved fludeoxyglucose F18. The emission computed tomography detection system acquires images of positron-emitting fludeoxyglucose F18 for the purpose of guiding the delivery of megavoltage X-rays for oncologic treatment with radiation therapy using an FDA cleared, authorized, or approved linear accelerator.
Reason for recall
Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potentially lead to adverse events (toxicity).
Data sourced from openFDA. This site is unofficial and independent of the FDA.