Recall Z-2485-2012— Biomet, Inc.

Class II · Terminated

Class II FDA medical device recall by Biomet, Inc., initiated 2012-08-23, terminated 2013-08-27. FDA's recall record names no specific 510(k) clearance for this event. Product: Optilock 2.7 mm Screw Inserter. Intended Use: The OptiLock Upper Extremity Plating System is intended for fixation of fractures and osteotomies. The System is intended for fractures and fracture dislocations, osteotomies and non-unions of the proximal humerus, particularly in osteopenic bone. The System is intended for fixation of fractures, osteotomies and non-unions of the olecranon, humerus, radius, ulna, particularly in osteopenic bone. The OptiLock VL Distal Radius Plating System is intended for use in distal radius and ulna (and other small bones), including use in osteopenic bone.. Reason: Field complaints indicate that recent shipments of the 14-400795 and 36505 drivers do not properly mate ("lock") into some of the specified screw heads. Drivers near MMC (Maximum Material Condition) or that are misshapen may not properly mate with screws at MMC. Design dimensions and tolerances of the drivers are not properly specified to prevent an incorrect fit between the drivers and the screw.

Recalling firm
Biomet, Inc.
Classification
Class II
Status
Terminated
Initiated
2012-08-23
Terminated
2013-08-27
Firm location
Warsaw, IN, United States

Product description

Optilock 2.7 mm Screw Inserter. Intended Use: The OptiLock Upper Extremity Plating System is intended for fixation of fractures and osteotomies. The System is intended for fractures and fracture dislocations, osteotomies and non-unions of the proximal humerus, particularly in osteopenic bone. The System is intended for fixation of fractures, osteotomies and non-unions of the olecranon, humerus, radius, ulna, particularly in osteopenic bone. The OptiLock VL Distal Radius Plating System is intended for use in distal radius and ulna (and other small bones), including use in osteopenic bone.

Reason for recall

Field complaints indicate that recent shipments of the 14-400795 and 36505 drivers do not properly mate ("lock") into some of the specified screw heads. Drivers near MMC (Maximum Material Condition) or that are misshapen may not properly mate with screws at MMC. Design dimensions and tolerances of the drivers are not properly specified to prevent an incorrect fit between the drivers and the screw

Data sourced from openFDA. This site is unofficial and independent of the FDA.