Recall Z-2482-2018— Cardinal Health 200, LLC

Class II · Terminated

Class II FDA medical device recall by Cardinal Health 200, LLC, initiated 2018-02-20, terminated 2019-11-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Presource PBDS, Kit, No Spiro. Kit includes a exerciser volumetric 5000 ML device, two items of cloth, skin prep 2% CHG, a packet of 4 chlorhexidine gluconate antiseptic cloths, and Gatorade pouches. Product Usage: The kit is provided to a patient pre-surgery, as part of the Enhanced Recovery After Surgery protocol. Items included at 1) an exerciser volumetric 5000ML device that is utilized to foster improved lung performance in an effort to avoid potential pulmonary post-operative complications, 2) up to two items of cloth, skin prep 2% CHG, a packet of for chlorhexidine gluconate antiseptic cloths utilized by the patient to reduce the risk of surgical site infections by wiping the body in a prescribed manner, 3) up to two Gatorade to help restore the body s electrolytes when fasting prior to surgery.. Reason: A defect in the seal of the Gatorade pouch contained in kits caused the pouch to leak within a limited portion of the kit..

Recalling firm
Cardinal Health 200, LLC
Classification
Class II
Status
Terminated
Initiated
2018-02-20
Terminated
2019-11-04
Firm location
Waukegan, IL, United States

Product description

Presource PBDS, Kit, No Spiro. Kit includes a exerciser volumetric 5000 ML device, two items of cloth, skin prep 2% CHG, a packet of 4 chlorhexidine gluconate antiseptic cloths, and Gatorade pouches. Product Usage: The kit is provided to a patient pre-surgery, as part of the Enhanced Recovery After Surgery protocol. Items included at 1) an exerciser volumetric 5000ML device that is utilized to foster improved lung performance in an effort to avoid potential pulmonary post-operative complications, 2) up to two items of cloth, skin prep 2% CHG, a packet of for chlorhexidine gluconate antiseptic cloths utilized by the patient to reduce the risk of surgical site infections by wiping the body in a prescribed manner, 3) up to two Gatorade to help restore the body s electrolytes when fasting prior to surgery.

Reason for recall

A defect in the seal of the Gatorade pouch contained in kits caused the pouch to leak within a limited portion of the kit.

Data sourced from openFDA. This site is unofficial and independent of the FDA.