Recall Z-2480-2012— Zimmer, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer, Inc., initiated 2011-04-01, terminated 2014-05-16. FDA's recall record names no specific 510(k) clearance for this event. Product: 00780302601 MIS 12/14 Slotted Femoral Head Provisional -3.5X26 00780302602 MIS 12/14 Slotted Femoral Head Provisional +0X26 00780302603 MIS 12/14 Slotted Femoral Head Provisional +3.5X26 00780302604 MIS 12/14 Slotted Femoral Head Provisional +7X26 00780302605 MIS 12/14 Slotted Femoral Head Provisional +10.5X26 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.. Reason: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp.

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Terminated
Initiated
2011-04-01
Terminated
2014-05-16
Firm location
Warsaw, IN, United States

Product description

00780302601 MIS 12/14 Slotted Femoral Head Provisional -3.5X26 00780302602 MIS 12/14 Slotted Femoral Head Provisional +0X26 00780302603 MIS 12/14 Slotted Femoral Head Provisional +3.5X26 00780302604 MIS 12/14 Slotted Femoral Head Provisional +7X26 00780302605 MIS 12/14 Slotted Femoral Head Provisional +10.5X26 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.

Reason for recall

Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp

Data sourced from openFDA. This site is unofficial and independent of the FDA.