Recall Z-2479-2018— Hitachi Medical Systems America Inc
Class II · Terminated
Class II FDA medical device recall by Hitachi Medical Systems America Inc, initiated 2018-06-18, terminated 2020-04-20. It names one FDA 510(k) clearance: K093044. Product: Hitachi Oasis MRI system Product Usage: Hitachi MR system is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The Oasis system produces transverse, coronal, sagittal, oblique, and curved crosssectional images that display the internal structure of the head, body, or extremities. The images produced by the MR systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.. Reason: The stainless steel belt which runs the length of the table underneath the tabletop is exposed when the tabletop is extended into the magnet for scanning. The firm has received two complaints where technologists have cut their fingers on this belt..
- Recalling firm
- Hitachi Medical Systems America Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-06-18
- Terminated
- 2020-04-20
- Firm location
- Twinsburg, OH, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Hitachi Oasis MRI system Product Usage: Hitachi MR system is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The Oasis system produces transverse, coronal, sagittal, oblique, and curved crosssectional images that display the internal structure of the head, body, or extremities. The images produced by the MR systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
Reason for recall
The stainless steel belt which runs the length of the table underneath the tabletop is exposed when the tabletop is extended into the magnet for scanning. The firm has received two complaints where technologists have cut their fingers on this belt.
Data sourced from openFDA. This site is unofficial and independent of the FDA.