Recall Z-2477-2012— Zimmer, Inc.
Class II · Terminated
Class II FDA medical device recall by Zimmer, Inc., initiated 2011-04-01, terminated 2014-05-16. FDA's recall record names no specific 510(k) clearance for this event. Product: 00789503601 VerSys Femoral Head Provisional 36mm -3.5 00789503602 VerSys Femoral Head Provisional 36mm +0 00789503603 VerSys Femoral Head Provisional 36mm +3.5 00789503604 VerSys Femoral Head Provisional 36mm +7 00789503605 VerSys Femoral Head Provisional 36mm +10.5 12/14 Neck Taper The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.. Reason: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp.
- Recalling firm
- Zimmer, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2011-04-01
- Terminated
- 2014-05-16
- Firm location
- Warsaw, IN, United States
Product description
00789503601 VerSys Femoral Head Provisional 36mm -3.5 00789503602 VerSys Femoral Head Provisional 36mm +0 00789503603 VerSys Femoral Head Provisional 36mm +3.5 00789503604 VerSys Femoral Head Provisional 36mm +7 00789503605 VerSys Femoral Head Provisional 36mm +10.5 12/14 Neck Taper The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
Reason for recall
Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
Data sourced from openFDA. This site is unofficial and independent of the FDA.