Recall Z-2476-2012— Zimmer, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer, Inc., initiated 2011-04-01, terminated 2014-05-16. FDA's recall record names no specific 510(k) clearance for this event. Product: 12/14 Neck Taper 00789503201 VerSys Femoral Head Provisional 32mm -3.5 00789503202 VerSys Femoral Head Provisional 32mm +0 00789503203 VerSys Femoral Head Provisional 32mm +3.5 00789503204 VerSys Femoral Head Provisional 32mm +7 00789503205 VerSys Femoral Head Provisional 32mm +10.5 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.. Reason: Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp.

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Terminated
Initiated
2011-04-01
Terminated
2014-05-16
Firm location
Warsaw, IN, United States

Product description

12/14 Neck Taper 00789503201 VerSys Femoral Head Provisional 32mm -3.5 00789503202 VerSys Femoral Head Provisional 32mm +0 00789503203 VerSys Femoral Head Provisional 32mm +3.5 00789503204 VerSys Femoral Head Provisional 32mm +7 00789503205 VerSys Femoral Head Provisional 32mm +10.5 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.

Reason for recall

Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp

Data sourced from openFDA. This site is unofficial and independent of the FDA.