Recall Z-2475-2019— Alere San Diego, Inc.
Class II · Terminated
Class II FDA medical device recall by Alere San Diego, Inc., initiated 2019-02-15, terminated 2023-12-12. It names one FDA 510(k) clearance: K120615. Product: Alere Cholestech LDX Analyzer, model # 14-874 containing power supplies with the identifying codes of 1837d and 1843d - Product Usage: The Alere Cholestech LDX System is a small, portable analyzer and test cassette system. The System is for in vitro diagnostic use only. The Lipid Profile GLU Cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides and glucose in whole blood. A TC/HDL (total cholesterol/HDL cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are calculated by the Alere Cholestech LDX¿ Analyzer.. Reason: The polarity of the output of power supply was reversed. Due to polarity of the output of power supplies being reversed, it renders the power supplies nonfunctional and the analyzer inoperable..
- Recalling firm
- Alere San Diego, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-02-15
- Terminated
- 2023-12-12
- Firm location
- San Diego, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Alere Cholestech LDX Analyzer, model # 14-874 containing power supplies with the identifying codes of 1837d and 1843d - Product Usage: The Alere Cholestech LDX System is a small, portable analyzer and test cassette system. The System is for in vitro diagnostic use only. The Lipid Profile GLU Cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides and glucose in whole blood. A TC/HDL (total cholesterol/HDL cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are calculated by the Alere Cholestech LDX¿ Analyzer.
Reason for recall
The polarity of the output of power supply was reversed. Due to polarity of the output of power supplies being reversed, it renders the power supplies nonfunctional and the analyzer inoperable.
Data sourced from openFDA. This site is unofficial and independent of the FDA.