Recall Z-2471-2023— Siemens Healthcare Diagnostics, Inc.

Class II · Ongoing

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2023-07-10. It names one FDA 510(k) clearance: K170065. Product: Atellica CH Total Bilirubin_2 -In vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma (lithium heparin) of adults and neonates Test Code: TBil_2 Siemens Material Number (SMN): 11097531. Reason: Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica¿ CH Total Bilirubin_2 (TBil_2), Atellica¿CH LDL Cholesterol (LDLC), Atellica¿ CH Gamma-Glutamyl Transferase (GGT) and Atellica¿CH HDL Cholesterol (HDLC).

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Ongoing
Initiated
2023-07-10
Firm location
Tarrytown, NY, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Atellica CH Total Bilirubin_2 -In vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma (lithium heparin) of adults and neonates Test Code: TBil_2 Siemens Material Number (SMN): 11097531

Reason for recall

Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica¿ CH Total Bilirubin_2 (TBil_2), Atellica¿CH LDL Cholesterol (LDLC), Atellica¿ CH Gamma-Glutamyl Transferase (GGT) and Atellica¿CH HDL Cholesterol (HDLC)

Data sourced from openFDA. This site is unofficial and independent of the FDA.