Recall Z-2468-2024— GE Medical Systems, LLC
Class II · Ongoing
Class II FDA medical device recall by GE Medical Systems, LLC, initiated 2024-06-14. It names 4 FDA 510(k) clearances: K163331, K160618, K142085 and 1 more. Product: Optima MR450w 1.5T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System. Reason: GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant..
- Recalling firm
- GE Medical Systems, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-06-14
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
4 on FDA's record- K163331 — Discovery MR750 3.0T +4 more
- K160618 — Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T, Optima MR450w 1.5T
- K142085 — DISCOVERY MR750 3.0T ,DISCOVERY MR450 1.5T, DISCOVERY MR750 3.0T,OPTIMA MR450W 1.5T
- K132376 — DISCOVERY MR750 3.0T, DISCOVERY MR450 1.5T, DISCOVERY MR750W 3.0T, OPTIMA MR450W 1.5T
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Optima MR450w 1.5T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
Reason for recall
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
Data sourced from openFDA. This site is unofficial and independent of the FDA.